Alarms
Resetting Alarms
TM80/TM70 Telemetry Monitor Operator’s Manual
6 - 9
The alarm pause time is two minutes. When the pause time expires, the alarm
paused status is automatically deactivated. You can also cancel the alarm paused
status by tapping [
Alarm Pause
] in the quick keys area.
6.9
Resetting Alarms
You can acknowledge the on-going alarms by resetting the alarms. After being reset, the
alarm reset symbol
is displayed in the message area.
When an alarm occurs, follow this procedure to reset the TM80/TM70’s alarm system.
■
Press the
key to enter the main menu, and then tap [
Alarm Reset
] from the
[
Commands
] section.
OR
1.
Flick your finger up at the bottom of the main screen to display the quick keys
area.
2.
Tap the [
Alarm Reset
] quick key to reset the alarm system.
6.9.1
Resetting Physiological Alarms
After the alarm system is reset, physiological alarms give the following alarm indicators:
■
The alarm sound is silenced.
■
The
symbol appears in the message area.
NOTE
•
When the TM80/TM70 is connected to the CMS and the function of
remotely pausing alarms is enabled at the CMS, alarms can be paused
either at the TM80/TM70 or at the CMS. For information on pausing
alarms at the CMS, refer to
BeneVision Central Monitoring System Opera-
tor’s Manual
.
NOTE
•
When the TM80/TM70 is connected to the CMS and the function of
remotely resetting alarms is enabled at the CMS, alarms can be reset
either at the TM80/TM70 or at the CMS. For information on resetting
alarms at the CMS, refer to
BeneVision Central Monitoring System Opera-
tor’s Manual
.