Alarms
Configuring the Alarms
TM80/TM70 Telemetry Monitor Operator’s Manual
6 - 7
6.7.2
Initiating Auto Alarm Limits
The TM80/TM70 provides the auto alarm limits function to automatically adjust alarm
limits according to the patient’s vital signs used. When auto limits are selected, the
TM80/TM70 calculates safe auto limits based on the latest measured values. To get accu-
rate auto alarm limits, you need to collect a set of measured vital signs as a baseline.
To initiate auto alarm limits, follow this procedure:
1.
Press the
key to enter the main menu.
2.
Tap [
Alarms
].
3.
Tap [
Alarm Limits
]. The [
Alarm Limits
] menu is displayed.
4.
Tap [
Auto Limits
].
5.
In the dialog that pops up, tap [
OK
].
Then the TM80/TM70 will automatically calculate alarm limits based on the latest mea-
sured values. Before applying these automatically created alarm limits, confirm if they
are appropriate for your patient from the Limits menu. If not, you can adjust them manu-
ally. These alarm limits will remain unchanged until you select auto limits again or adjust
them manually.
The TM80/TM70 calculates auto limits basing on the following rules:
NOTE
•
When the TM80/TM70 is connected to the CMS, any changes to the alarm
properties at the TM80/TM70 or the CMS will be communicated to the
other side to make sure that the alarm settings are consistent.
Module
Parameter
Lower Limit
Upper Limit
Auto Limit Range
Adult/
Pediatric
Adult/
Pediatric
ECG
HR/PR
(bpm)
(HR/PR×0.8) or
40 (whichever is
greater)
(HR/PR×1.25) or
240 (whichever
is smaller)
Adult/pediatric:
35~240
Resp
RR
(rpm)
(RR×0.5) or 6
(whichever is
greater)
(RR×1.5) or
30(whichever is
smaller)
Adult/pediatric:
6~55
SpO
2
SpO
2
(%)
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
Summary of Contents for BeneVision TM80
Page 2: ......