Pausing Alarms
Alarms
6 - 8
TM80/TM70 Telemetry Monitor Operator’s Manual
6.7.3
Restoring the Default Alarm Settings
1.
Press the
key to enter the main menu.
2.
Tap [
Alarms
].
3.
Tap [
Alarm Limits
]. The [
Alarm Limits
] menu is displayed.
4.
Tap [
Defaults
].
5.
In the dialog that pops up, tap [
OK
].
6.8
Pausing Alarms
When an alarm occurs, follow this procedure to pause the alarm system.
■
Press the
key to enter the main menu, and then tap [
Alarm Pause
] from the
[
Commands
] section.
OR
1.
Flick your finger up at the bottom of the main screen to display the quick keys
area.
2.
Tap the [
Alarm Pause
] quick key.
When alarms are paused, the following indicators are given:
■
No physiological alarm will be presented within the alarm pause time.
■
For technical alarms, alarm sounds are paused, but the alarm lamp and alarm
messages remain presented.
■
The remaining alarm pause time is displayed in the message area.
■
The alarm pause symbol
is displayed in the upper right corner of the main
screen.
NIBP
NIBP-S
(mmHg)
(SYS×0.68
+
10)
SYS×0.86
+
38
Adult: 45~270
Pediatric: 45~185
NIBP-D
(mmHg)
(Dia×0.68
+
6)
(Dia×0.86
+
32)
Adult: 25~225
Pediatric: 25~150
NIBP-M
(mmHg)
(Mean×0.68
+
8)
Mean×0.86
+
35
Adult: 30~245
Pediatric: 30~165
Module
Parameter
Lower Limit
Upper Limit
Auto Limit Range
Adult/
Pediatric
Adult/
Pediatric
Summary of Contents for BeneVision TM80