Preparation for Monitoring ECG
Monitoring ECG
7 - 8
TM80/TM70 Telemetry Monitor Operator’s Manual
7.3.4.4
Lead Placement: Pacemaker Patients
The recommended lead placement for monitoring a pacemaker patient is as follows.
LA (black)
L (yellow)
Under the patient’s left
clavicle, at the mid-clavicular
line within the rib cage frame
LL (red)
F (green)
On the patient’s lower left
abdomen within the rib cage
frame.
RL (green)
N (black)
On the patient’s lower right
abdomen within the rib cage
frame.
Va
(brown)
Ca
(white)
The Va (Ca) electrode is
placed in any one of the posi-
tion from V1 (C1) to V6 (C6).
By default, the Va electrode is
placed at V1 while the Ca
electrode is placed at C1.
V1 (brown)
C1 (white)
In the fourth intercostal
space, right sternal border.
V2 (brown)
C2 (white)
In the fourth intercostal
space, left sternal border.
V3 (brown)
C3 (white)
In the midway between V2
and V4 on a straight line.
V4 (brown)
C4 (white)
In the fifth intercostal space,
mid-clavicular line.
V5 (brown)
C5 (white)
In the fifth intercostal space,
anterior axillary line.
V6 (brown)
C6 (white)
In the fifth intercostal space,
mid-axillary line.
AHA
IEC
Electrode Placement