Page 48
MERCAP – INSTRUCTION MANUAL
7ML19981CM01.1
Command #130 Set Upper Sensor Limit
This command sets the Upper Sensor Limit value to the actual P.V. value.
Request data bytes
None
Response data bytes
None
Command #131 Set Lower Sensor Limit
This command sets the Lower Sensor Limit value to the actual P.V. value.
Request data bytes
None
Response data bytes
None
Command #132 Write Sensor Limit Values
This command writes specific values to the Upper and Lower Sensor Limits. The Units
selection is only accepted in pF. The Lower Sensor Limit Value must not be less than zero and
not more than the Upper Sensor Limit Value.
The Upper Sensor Limit Value may not be more than 3300 and no less than the Lower Sensor
Limit. The minimum distance between Upper- and Lower Sensor Limit is forced to be at least
3.3 pF. The actual Upper- and Lower Sensor Limit Values can be read with command #14
Read Primary Variable Sensor Information.
Request data bytes
Data byte
#0
Sensor
Limits Units
#1
Upper
Sensor Limit
MSB
#2
#3
#4
Upper
Sensor Limit
LSB
#5
Lower
Sensor Limit
MSB
#6
#7
#8
Lower
Sensor Limit
LSB
Data byte #0
Sensor Limits Units Code, 8-bit unsigned integer, must
be 153 (pF)
Data byte #1-#4
Sensor Upper Limit Value, IEE754
Data byte #5-#8
Sensor Lower Limit Value, IEE754