MERCAP – INSTRUCTION MANUAL
7ML19981CM01.1
HART
TM
Response Code information
Additional response code information, Second Byte.
Bit #7: Field Device Malfunction
When the transmitter detects a malfunction, the Analog Output will be set in a fault state.
Bit #6: Configuration Changed
When any of the settings in EEROM is changed either by a write command or by manual
ZERO or SPAN adjust, this bit is set. Use command 38 to reset.
Bit #5: Cold Start
This bit is issued once after an initialization cycle is complete; this can occur after a power
loss or as a result of a (watchdog) reset.
Bit #4: Extended Status Available
When any of the extended status bits is set this flag is raised. Use command 48 to get
detailed status information.
Bit #3: Output Current Fixed
This bit is set as long as the Primary Variable Analog Output is set to a fixed value.
Bit #2: Primary Variable Analog Output Saturated
Flag is set when the Primary Analog Output saturates below 3.8 mA and above 20.5 mA.
Bit #0: Primary Variable Out of Limits
This flag is set whenever the Transmitter Variable #0 (in pF), the Primary Variable exceeds the
Sensor Limits returned with Command 14, Read Primary Variable Sensor Limits.
General transmitter information
Damping information
The MST9500 transmitter implements damping only on the Analog Output Current signal. This
is a fixed algorithm.
Non-volatile Memory Data Storage
The flags byte of Command #0 referenced in the Universal Command Specification document,
will have Bit #1 (Command #39, EEPROM Control Required) set to 0, indicating that all data