The possible checks can be subdivided in two main groups:
1.
Visual Checks
a.
inside enclosure clean and dry
b.
enclosure sealing intact and working properly (not hardened)
c.
all screw connections are tight
d.
ground connections inside intact
e.
ground connections outside intact
f.
no oxidation on push-button and 15 pole source connector
g.
no dirt or deposits on coax connector
h.
no cable or wires jammed under cover
2.
Functional Checks
a.
provides manual test function 0.25 mA current increase during 10 seconds
b.
check for required minimum terminal voltage
c.
does the current go to the alarm position (3.6 or 22 mA) if the coax plug is
unplugged? If so, fasten it again.
d.
via HART
TM
Does the PV go to 0 pF when the coax plug is unplugged (
±
0.15 pF is allowed)? If
so, switch the output current to 4 respectively 20 mA and check the current through
the loop.