Note:
Example to explain the burst interval in conjunction with the ext. warmup time (ext. warmup time <
measurement cycle): warmup time (ext. warmup time < measurement cycle):
Basic setting
Record interval
5min.
Burst interval
60 sec.
Measurement cycle
15 sec.
Output channels
Ext warmup time
5
Unit
sec.
Measurement channels -
>Config.
Decay
med
Time
60 sec.
Output on the device
Sensor supply
Formation of the measurement value:
In each case, the sensor supply is activated 5 sec before expiry of the
measurement cycle. This results in four valid measurements that are used to determine a median and are
recorded as the measurement value.