Page 42
Malfunctions, Troubleshootings
optoNCDT ILR 1183
9.
Malfunctions, Troubleshootings
9.1
Malfunctions
Defaults
Cause
Action for removal
No data coming via Profibus Faulty Profibus configuration
Check Profibus configuration
Device error
(Ext. diagnosis)
Hardware problems
Reship sensor for repair, contact technical
support
9.2
Troubleshootings via Profibus
Code Cause
Action for removal
E15
Excessively poor reflexes
Use target board, observe minimum requirement
on measuring distance (> 0.1 m)
E16
Excessively strong reflexes
Use target board
do not measure against reflecting surfaces
E17
Too much steady light
(for example sun)
Mount sensor in such a position that excessive incidence of
steady light is prevented, extend glare protection tube, pro-
vide additional light-shielding, for example protective cap
E18
Only in DX mode: too much difference
between measured and pre-calculat-
ed value
Check path from distance meter to target being measured
for obstacles
E23
Temperature below –10 °C/ +14 °F
Provide ambient temperature > -10 °C/ +14 °F
E24
Temperature above +60 °C/ +140 °F Provide ambient temperature < +47 °C/ +116.6 °F
E31
Faulty EEPROM checksum, hardware
error
Service required if fault occurs repeatedly
> Reship sensor for repair
E51
Failure to set avalanche voltage
1. straylight
2. hardware error
1. Check target reflectance and ambient light
(steady light), make sure that neither target not entry open-
ing of sensor is exposed to light from reflecting surfaces,
projectors or sun
2. Service required > reship for repair