Page 26
Installation
optoNCDT ILR 1183
The following parameters are available for configuration (refer to GSD File):
- ExtUserPrmData = 20 “Trigger Mode” for trigger mode
- ExtUserPrmData = 21 “Trigger Level” for trigger level
- ExtUserPrmData = 25 “Trigger Delay (31-16)” for trigger delay
- ExtUserPrmData = 26 “Trigger Delay (15-0)” for trigger delay
For detection of a clock edge, the following voltage signals are required:
- 24 V > HIGH > 11 V
- 0 V < LOW < 6.5 V
Wiring connections for working with the trigger input are as follows:
A
B
Pin 2
Pin 4
A
B
Pin 2
Pin 4
Profibus OUT
Profibus IN
GND
U
v
violet
red
green
supply
application
TRIG
Fig. 12 Wiring of trigger input