Page 13
Functional Principle, Technical Data
optoNCDT ILR 1183
Model
ILR 1183
Switching outputs
Open Collector, HIGH = U
V
– 2 V / LOW < 2 V (max 500 mA), short-circuit-proof; switching
threshold and hysteresis selectable
Trigger input
Trigger edge and delay selectable, trigger pulse of max. 24 V
Serial interface
SSI Interface (RS422), 24 bit, gray-coded,
Transfer rate: 50 kHz ... 1 MHz, 200 μs break
Profibus data interface
3
Profibus RS485 DP-V0 slave under IEC 61158 / IEC 61784,
Transfer rate: 9.6 kbaud … 12 MBaud, identno. 0x09CB,
ILR809CB.GSD, PNO Profile Encoder Class 1/2, configuration of measuring parameters, output
of measured values and error messages,
parameters and PB-address are stored in NVRAM
Operating mode
Individual measurement, external trigger, distance tracking, continuous
measurement
Analog output
-
Trigger input
One input with HIGH signal level > 11 V and LOW signal level < 6.5 V
2.5 mA input current at 24 V, trigger edge and delay selectable,
trigger pulse of max. 24 V
Power supply
10 … 30 VDC
Max. power consumption
< 3.2 W at 24 VDC
Connection
1 x 12-pole (Binder series 723) M16, 2 x 5-pole (Binder series 766) M12 B-coded
Protection class
IP 65
Dimensions
210 mm x 99 mm x 51 mm
Housing material
Extruded aluminum profile with powder-coat paint finish
Weight
980 g
1) Conditional on target reflectance, ambient light influences and atmospheric conditions
2) Statistic variation 95 %
3) Sensor settings occur about this interface.