■■■■■■■■■■■■■■■■■■■■■■
6 Troubleshooting
873 Biodiesel Rancimat
■■■■■■■■
161
Problem
Cause
Remedy
The induction times are
not reproducible for
multiple determina-
tions.
The reaction vessels used are
not clean.
■
Free the reaction vessel of particles
(dust, cardboard, glass fragments etc.)
with compressed air before weighing
in the sample.
■
Only use new, unused reaction ves-
sels.
The reaction vessels used are
scratched on the inside.
Only use new, unused reaction vessels.
The reaction vessel cover
(5-
) is not mounted correctly
or is too loose.
■
If the reaction vessel cover is oblique
or not completely mounted, press
down as far as the limit stop.
■
If the reaction vessel cover is loose on
the reaction vessel despite correct
assembly, the cover has to be
replaced.
The connection to the measur-
ing vessel is not mounted cor-
rectly.
Ensure that no air can escape through
leaks when transferring from the reaction
vessel to the measuring vessel.
The temperature in different
channels of a heating block
differs, as there are burn marks
at one or more places in the
opening of the heating block.
Check temperature in the different posi-
tions
(see Chapter 4.8.4.1, page 141)
. If
necessary, carefully remove the impurities
from the
cold
heating block.
The temperature in different
heating blocks differs.
Determine the temperature correction
Delta T for each block separately
.
The sample is not homoge-
nous.
Homogenize sample.
The stability times are
not reproducible for
multiple determina-
tions.
The cell constant was not
determined or does not corre-
spond to the value entered.
■
Determine cell constant
.
■
Ensure that the measuring cell is
always used in the same channel, so
that the cell constant determined
once also corresponds to the measur-
ing cell.
■
Ensure that the measuring cell is not
contaminated. If necessary, clean
.
Summary of Contents for 1.873.0014
Page 1: ...873 Biodiesel Rancimat Manual 8 873 8003EN...
Page 2: ......