4.8 GLP functions
■■■■■■■■■■■■■■■■■■■■■■
148
■■■■■■■■
873 Biodiesel Rancimat
■
Click on the menu item
Tools
▶
GLP test
▶
Conductivity...
and
confirm with
[Next >]
.
The following dialog window appears:
■
Select the option
Use standard solution
.
■
Under
Conductivity of standard solution
, enter the conductivity
in µS/cm and confirm with
[Next >]
.
c(KCl) = 1 mmol/L
can be used as a standard solution. This is derived
from the optionally available conductivity standard 6.2301.060
(KCl 0.1 mol/L) through dilution with distilled water. The conductivity
of this diluted solution is:
Temperature Conductivity
18 °C
19 °C
20 °C
21 °C
22 °C
23 °C
24 °C
25 °C
127 µS/cm
130 µS/cm
133 µS/cm
136 µS/cm
138 µS/cm
141 µS/cm
144 µS/cm
147 µS/cm
The conductivity is measured sequentially at all channels.
After the last measurement is completed, the dialog window with the
test results
appears.