3.3 Mains connection
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
873 Biodiesel Rancimat
3.3.2
Replacing fuses
Two fuses 4 A (slow-acting) for 115 V or 2 A (slow-acting) for 230 V are
installed in the fuse holder
of the 873 Biodiesel Rancimat as standard.
Warning
Ensure that the instrument is never put into operation with fuses of
another type, otherwise there is a risk of fire!
Proceed as follows to replace defective fuses:
1 Pull out the mains cable
■
Pull the mains cable out of the mains connection socket of the
Biodiesel Rancimat.
2 Remove the fuse holder
■
Use a screwdriver to loosen and completely pull out the fuse holder
above the mains connection socket.
3 Replace fuses
■
Carefully take the defective fuses out of the fuse holder and replace
them with two new fuses suitable for the set mains voltage of the
type TH (slow-acting, with high switching capacity):
– 115 V 4 A (TH) Metrohm No.: U.600.0022
– 230 V 2 A (TH) Metrohm No. U.600.0107
4 Insert the fuse holder
■
Push the fuse holder back into the instrument until it latches into
place.
3.3.3
Mains cable and mains connection
Mains cable
The mains cable optionally supplied for the instrument
■
6.2122.020 with plug SEV 12 (Switzerland)
■
6.2122.040 with plug CEE(7), VII (Germany, …)
■
6.2122.070 with plug NEMA 5-15 (USA, …)
is three-core and provided with a plug with grounding pin. If another plug
has to be mounted, the yellow/green conductor (IEC standard) must be
connected to the protective ground (protection class I).