4 Operation
873 Biodiesel Rancimat
■■■■■■■■
145
5 Carry out measurement for further channels
Warning
The aluminum cylinder of the GLP test set is very hot. Do not touch!
■
As soon as the message appears with the prompt to insert the
external temperature sensor together with the measuring insert in
the next selected channel,
rapidly carry out the change
.
■
Carry out measurement for each of the selected channels.
After the last measurement is completed, the dialog window with the
test results
appears.
6 Confirm or reject results
✓
Test passed
: The deviation of the measured temperatures
is within the specified tolerance.
✗
Test failed
: The deviation of the measured temperatures is
greater than the specified tolerance.
■
Confirm the results with
[Finish]
.
or
■
Reject the results with
[Cancel]
.