4.4 Calibration functions
■■■■■■■■■■■■■■■■■■■■■■
52
■■■■■■■■
873 Biodiesel Rancimat
Preparing Delta T determination
shows in detail how the accessories for determining Delta T are
mounted. Proceed as follows:
1 Prepare reaction vessel cover
■
Mount the air tube on the reaction vessel cover
reaction vessel cover", page 15)
.
■
Clamp the first spacer at a distance of approx. 12 cm from the lower
end onto the air tube.
■
Clamp the second spacer at a distance of approx. 5 cm from the
lower end onto the air tube.
■
Insert the temperature sensor from above into the temperature
sensor opening of the reaction vessel cover and fasten it in the cor-
responding openings of the spacers.
2 Prepare reaction vessel
■
Fill the reaction vessel with 5 g silicone oil (e.g. FLUKA 85409).
■
Place the reaction vessel cover with the temperature sensor on the
reaction vessel.
■
Push the temperature sensor fully downwards (the sensor must be
on the vessel base).
3 Insert and connect reaction vessel
■
Insert the reaction vessel with the mounted reaction vessel cover in
channel 2 or 3 of the required heating block.
■
Screw the one end of the FEP tubing 250 mm onto the thread
adapter M8 / M6 of the reaction vessel cover.
■
Screw the other end of the FEP tubing onto the corresponding con-
nector
4 Set method parameters
Note
Create a unique method for both blocks, so that you can separately
enter the temperature correction for both blocks.
■
In the
Control window
, select the required
method
for the heat-
ing block for which the temperature correction Delta T is to be
determined