4.4 Calibration functions
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
873 Biodiesel Rancimat
Time
Time of the event. The format depends on the settings defined in Windows
in
Control Panel
▶
Country settings
▶
Time
.
User
Name of the user logged in at the time of the event.
Unit
Instrument number (1…4).
Event
Description of the event.
4.4
Calibration functions
4.4.1
Determining cell constants
Only the
change in conductivity
is measured an evaluated during the
Rancimat measurement. As a result, the cell constants of the conductivity
measuring cells used in the Biodiesel Rancimat, which are normally in the
range from 1.00...1.20, only have to be entered exactly if the
absolute
conductivity measured value
is to be displayed correctly.
The value also has to be entered before carrying out the
GLP test "Con-
ductivity"
. The cell constants can thereby be entered manually or deter-
mined automatically using a defined standard solution
Entering cell constants manually
Proceed as follows if you wish to enter the cell constants manually:
1 Open dialog window
■
Click on the menu item
Tools
▶
Cell constants
▶
Manual...
.
The following dialog window appears: