4.4 Calibration functions
■■■■■■■■■■■■■■■■■■■■■■
50
■■■■■■■■
873 Biodiesel Rancimat
■
Enter the conductivity of the standard used.
c(KCl) = 1 mmol/L
can be used as a standard solution. This is derived
from the optionally available conductivity standard 6.2301.060
(KCl 0.1 mol/L) through dilution with distilled water. The conductivity
of this diluted solution is:
Temperature Conductivity
18 °C
19 °C
20 °C
21 °C
22 °C
23 °C
24 °C
25 °C
127 µS/cm
130 µS/cm
133 µS/cm
136 µS/cm
138 µS/cm
141 µS/cm
144 µS/cm
147 µS/cm
Alternatively, the conductivity standard 6.2324.000 (undiluted) can be
used.
3 Start determination and check the values
■
Click on
[Start]
to start the determination of the cell constant.
After the determination is finished, the dialog window
Enter/Check
cell constants
appears again.
If you wish to accept the displayed values, click
[OK]
.
If you wish to maintain the old values, click
[Cancel]
.
4.4.2
Determining Delta T
The
temperature correction Delta T
indicates the deviation of the current
temperature of the sample from the temperature in the heating block and
belongs to the method as a parameter
can automatically be determined with the calibrated, external temperature
sensor (part of the optionally available GLP test set,