4.6 Determinations
■■■■■■■■■■■■■■■■■■■■■■
80
■■■■■■■■
873 Biodiesel Rancimat
The determination is automatically saved in the database
Repos.mrd
. The channel is ready for starting a new determination.
2 Clean vessels and accessories
Note
Always clean the vessels and accessories
directly after each deter-
mination
4.6.4
Cleaning the instrument and accessories
1 Clean the measuring vessels and accessories
■
Clean used
measuring vessels
after pouring off the measuring
solution with acetone or 2-propanol. Pre-clean with washing-up
liquid in the case of severe contamination.
■
Rinse the product thoroughly with distilled water.
■
Clean the
measuring vessel covers
PTFE tubes
electrodes
with acetone or 2-propanol and
rinse with distilled water. Pre-clean with washing-up liquid in the
case of severe contamination.
Remove the protective ring
to be able to clean the electrodes.
2 Clean the reaction vessels and accessories
■
Dispose of used
reaction vessels
and
air tubes
and use new
reaction vessels and air tubes for the next measurement.
■
Clean the
reaction vessel covers
with acetone or 2-propanol and
rinse with distilled water. Pre-clean with washing-up liquid in the
case of severe contamination.
■
Then heat the
reaction vessel covers
in the drying oven for 2
hours at 80 °C.
Note
Replace the reaction vessel covers if they are not tightly on the reac-
tion vessel anymore or the material is obdurate.
3 Clean the tubings
■
Clean the
Iso-Versinic
®
tubings
and rinse with distilled water. Pre-clean with washing-up liquid in
the case of severe contamination.