MI 3205 TeraOhm XA 5kV
Instrument description
12
2.3 Standards applied
The TeraOhm XA 5kV instrument is manufactured and tested in accordance with the
following regulations:
Electromagnetic compatibility (EMC)
EN 61326
Electrical equipment for measurement, control and laboratory
use – EMC requirements Class A
Safety (LVD)
EN 61010 - 1
Safety requirements for electrical equipment for measurement,
control and laboratory use – Part 1: General requirements
EN 61010 - 2 - 030
Safety requirements for electrical equipment for measurement,
control and laboratory use – Part 2-030: Particular requirements
for testing and measuring circuits
EN 61010 - 031
Safety requirements for hand-held probe assemblies for electrical
measurement and test.
Some further recommendations
IEEE 43 – 2000
Recommended Practice for Testing Insulation Resistance of Rotating
Machinery:
1 M
Ω
+ 1 M
Ω
/ 1000 V rating of equipment for insulation systems prior to 1970;
5 M
Ω
for random wound motors under 600 Volts;
100 M
Ω
for form wound motors over 600 Volts, and armatures;
IEC 60439-1
Low-voltage switchgear and control gear assemblies - Part 1:
Type-tested and partially type-tested assemblies:
Scope of the insulation resistance test: alternative method for verification of
dielectric properties by insulation resistance measurement;
Description: a d.c. test voltage (500 V) is applied to the insulation and its
resistance is measured. Insulation is proper, if its resistance is high enough (1000
Ω
/ V of circuit rated voltage);
IEC 61558
Isolating transformers and safety isolating transformers:
Test voltage: 500 V, measurement period: 1 min;
Minimum insulation resistance for basic insulation: 2 M
Ω
;
Minimum insulation resistance for supplementary insulation: 5 M
Ω
;
Minimum insulation resistance for reinforced insulation: 7 M
Ω
;
Li – ion battery pack
IEC 62133
Secondary cells and batteries containing alkaline or other non-
acid electrolytes - Safety requirements for portable sealed
secondary cells, and for batteries made from them, for use in
portable applications.
Note:
Immunity to radiated RF fields (Field strength: 10V/m, Modulation: AM, 80%, 1 kHz)
Voltage Range
Operational conditions
Disturbance < 5 %
Disturbance > 5 %
50 V
100 M
Ω
300 MHz ÷ 900 MHz
900 MHz ÷ 1 GHz
Note about EN and IEC standards:
Text of this manual contains references to European standards. All standards of
EN 6XXXX (e.g. EN 61010) series are equivalent to IEC standards with the same
number (e.g. IEC 61010) and differ only in amended parts required by European
harmonization procedure.