MI 3205 TeraOhm XA 5kV
Technical specifications
56
Current
Measuring range (A)
Resolution (A)
Accuracy
1.00
5.00 m
10
(5 % of r 3 digits)
100
999
1
10.0
99.9
100 n
1.00
9.99
10 n
100
999 n
1 n
10.0
99.9 n
100 p
0.00
9.99 n
10 p
(10 % of r 0.15 nA)
Table 8.2: Current measurement ranges and accuracy
Voltage
Measuring range (V)
Resolution (V)
Accuracy
0
999
1
(5 % of r 3 digits)
1.00
5.99 k
10
Table 8.3: Voltage measurement ranges and accuracy
Capacitance
Measuring range (F)
Resolution (F)
Accuracy
20
999 n
1 n
(5 % of r 3 digits)
1.00
9.99
10 n
10.0
50.0
100 n
Table 8.4: Capacitance measurement ranges and accuracy
Nominal voltage range ................................... 500 V
5 kV
Dielectric absorption ratio DAR
Display range DAR
Resolution
Accuracy
0.01
9.99
0.01
(5 % of r 3 digits)
10.0
100.0
0.1
Table 8.5: DAR display ranges and accuracy
Polarization index PI
Display range PI
Resolution
Accuracy
0.01
9.99
0.01
(5 % of r 2 digits)
10.0
100.0
0.1
Table 8.6: PI display ranges and accuracy