MI 3205 TeraOhm XA 5kV
Main menu
29
4.4.9 Break Down selection
In this selection the Break down can be set.
When triggered the Break down circuit
automatically stops the measuring procedure.
Keys:
Toggles between YES or NO (changes are stored automatically).
Notes:
Break down is not active in Withstanding Voltage test and in the charging state of
the high voltage generator.
Break down is active if output voltages higher or equal then 1kV are set!
4.4.10 Instrument
info
In this menu the following instrument data is shown:
firmware version;
serial number;
calibration date;
battery type.
4.5 Help Menu
The Help menu contains schematic diagrams for illustrating how to properly connect the
instrument to the various test objects.
Keys in help menu:
Selects next / previous help screen.
ESC
Returns to the
Settings menu
.
Figure 4.11: Examples of help screens