MI 3205 TeraOhm XA 5kV
Instrument description
18
Timer2
Delay for starting PI measurement (mm:ss) – step 1 s (max time
99 min).
Timer3
Allows user to set the desired measurement duration (mm:ss) –
step 1 s (max time 99 min).
DD
Allows user to enable or disable Dielectric Discharge
measurement.
Itrgg
Allows user to set the desired trigger level – step 100
μ
A (max
current 5 mA).
Tstart
Allows user to set the start test voltage time (mm:ss) – step 1s
(max time 99 min).
Tend
Allows user to set the constant end test voltage time (mm:ss) –
step 1 s (max time 99 min).
Ustart
Allows user to set the desired start test voltage.
Uend
Allows user to set the desired end test voltage.
Tramp
Allows user to set the duration of the test ramp (mm:ss) – step 1
s (max time 99 min).
HI Lim
Allows user to set the high limit (value is evaluated at the end of
the measurement).
AVG
Allows user to set the additional averaging of the result (OFF, 5,
10, 30, 60).
3.4.3 Message window
In the message field warnings and messages are displayed.
Figure 3.7: Message window