5.6 Insulation Resistance Measurement
Test can be started from the Insulation Resistance Measurement window. Before
carrying out a test the parameters output voltage, timer, high limit and additional
averaging can be edited.
Figure 5.8: Insulation Resistance Measurement menu
Test parameters for Insulation resistance Measurement
Un
Set test voltage – step 50 V (50 V – 1 kV) and 100 V (1 kV – 5 kV).
Timer1
Duration of measurement (mm:ss) – step 1 s (max time 99 min).
Hi Lim
Limit value selection (OFF, 0.50 M
Ω
– 1.0 T
Ω
).
AVG
Additional averaging of the result value (OFF, 5, 10, 30, 60).
Keys:
Selects the field to be changed.
Modifies selected field.
SELECT
Toggles between graph view and results view. (Graph must be
enabled in the Settings menu).
MEM
Enters Save custom test menu.
Saves results (if present).
START/STOP
Starts or Stops the Insulation Resistance Measurement.
ESC
Exits to Measurements menu.
Keys in graph screen – measurement completed:
Scroll the cursor along displayed graph data.
Cursors on / off