MI 2893 / MI 2892 / MI 2885
Periodic calibration
241
7.6
Periodic calibration
To ensure correct measurement, it is essential that the instrument is regularly calibrated. If used
continuously on a daily basis, a six-month calibration period is recommended, otherwise annual
calibration is sufficient.
7.7
Service
For repairs under or out of warranty please contact your distributor for further information.
7.8
Troubleshooting
If ESC button is pressed while switching on the instrument, the instrument will not start. Batteries have
to be removed and inserted back. After that the instrument will start normally.