MI 2893 / MI 2892 / MI 2885
Measurements
219
Note: Overall current accuracy (as percent of measured value), is provided as guideline. For exact
measuring range and accuracy please check user manual of related current clamps. Overall accuracy is
calculated as:
2
2
acy
ClampAccur
Accuracy
Instrument
1,15
uracy
OverallAcc
Half cycle RMS current (inrush, min, max) I
1Rms(1/2)
, I
2Rms(1/2)
, I
3Rms(1/2)
, I
NRms(1/2)
, AC+DC
Clamps
Range
Measuring range
Overall current accuracy
A 1281
1000 A
100 A
5 A
0.5 A
100 A ÷ 1200 A
10 A ÷ 175 A
0.5 A ÷ 10 A
50 mA ÷ 1 A
±0.8 %
I
RMS
A 1588
50 A
5 A
0.5 A
5 A ÷ 100 A
0.5 A ÷ 10 A
50 mA ÷ 1 A
±0.8 %
I
RMS
A 1398 PQA
10 A
0.5 A ÷ 20 A
±0.8 %
I
RMS
A 1033
1000 A
100 A
20 A ÷ 1000 A
2 A ÷ 100 A
±1.3 %
I
RMS
A 1069
100 A
10 A
5 A ÷ 200 A
500 mA ÷ 20 A
±1.3 %
I
RMS
A 1391 PQA
100 A
10 A
5 A ÷ 200 A
500 mA ÷ 20 A
±1.5 %
I
RMS
A 1636
DC: 2000 A
AC: 1000 A
40 A ÷ 2000 A
20 A ÷ 1000 A
±1.5 %
I
RMS
A 1227
3000 A
300 A
30 A
300 A ÷ 6000 A
30 A ÷ 600 A
3 A ÷ 60 A
±1.6 %
I
RMS
A 1227 5M
3000 A
300 A
30 A
300 A ÷ 6000 A
30 A ÷ 600 A
3 A ÷ 60 A
±1.6 %
I
RMS
A 1445
3000 A
300 A
30 A
300 A ÷ 6000 A
30 A ÷ 600 A
3 A ÷ 60 A
±1.6 %
I
RMS
A 1582
3000 A
300 A
30 A
300 A ÷ 6000 A
30 A ÷ 600 A
3 A ÷ 60 A
±1.6 %
I
RMS
A 1501
3000 A
300 A
30 A
300 A ÷ 6000 A
30 A ÷ 600 A
3 A ÷ 60 A
±1.6 %
I
RMS
A 1502
3000 A
300 A
30 A
300 A ÷ 6000 A
30 A ÷ 600 A
3 A ÷ 60 A
±1.6 %
I
RMS
A 1503
6000 A
600 A
60 A
600 A ÷ 12 000 A
60 A ÷ 1200 A
6 A ÷ 120 A
±1.6 %
I
RMS
A 1446
6000 A
600 A
60 A
600 A ÷ 12 000 A
60 A ÷ 1200 A
6 A ÷ 120 A
±1.6 %
I
RMS
A 1037
6 A
0.5 A
0.5 A ÷ 10 A
10 mA ÷ 10 A
±0.4 %
I
RMS