MI 2893 / MI 2892 / MI 2885
E-Meter recorder (MI 2892/MI 2885)
102
Note: Recorder will run until one of the following end conditions is
met:
STOP key was pressed by user
Given Duration criteria was met
SD CARD is full
Note: If during record session instrument batteries are drained, due
to long interruption for example, instrument will shut down
automatically. After power restauration, it will automatically start
new recording session.
Measurements number
The n-th measurement is performed out of the total defined number
of measurements
Example: 3/5
the third measurement is active from the total
number of 5.
Registered energy
Energy (kWh/kvarh) registered by the Power Quality Instrument
Number of pulses
Number of accepted pulses from the E-Meter from the total defined
number of pulses.
Example: 3/5
3 pulses accepted from the total selected number
of 5.
Error
E-Meter error, getting from the individual measurement. On the LCD
screen only last 5 error results are presented. For the detailed view,
download data with PowerView and analyse the data.
Average error
Average error calculated from the individual measurements.
Table 82: Functional Keys in E-Meter recorder setup screen
START
STOP
Starts the recorder (selection, when recorder is not running)
Stops the recorder (selection, when recorder is running)
RESET
Reset E-Meter measurements results
(selection, when recorder is not running)
CONFIG
Shortcut to the PQI Connection setup.
(Selection, when recorder is not running)
SETUP
E-Meter recorder setup.
(Selection, when recorder is not running)
Moves cursor to the next row up/down; enter desired
character
Field selection/Increase (Decrease) the value
F1
F2
F3
F4