EX1200A Triggering
75
Indicators section of Section 3).
F
IGURE
4-1:
I
NPUT
P
IN
L
OCATIONS AND
S
IGNALS
Standard banana plugs can be used to interface other external instruments with the DMM. The
fuse located on the front panel is for current input protection and is user-replaceable. It is a 3.15 A,
250 V fuse and can be replaced with VTI P/N: 35-0054-315.
NOTE
Use sheathed-type probes for voltage and current measurements at the front panel of DMM.
Rear Panel Connector
Mainframes with an EX1200A DMM installed provide an onboard DMM with a five-line analog
bus routed through the backplane to the DMM. This bus is available through rear panel DMM Bus
connector that can be used to extend the analog backplane to other EX1200A series mainframes.
To ensure high signal integrity, environmental noise and cable length should be accounted for
prior to routing these lines.
CAUTION
When voltages are switched across the backplane using a multiplexer or matrix switch card, these
voltages are present on the DMM Bus connector. Caution must be observed when handling this
connector.
Also, when the DMM bus is extended to other EX1200A mainframes, the user must ensure that no
two devices are using the bus simultaneously.
Summary of Contents for EX1200A SERIES
Page 32: ......
Page 50: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use...
Page 63: ...www vtiinstruments com EX1200A Series Preparation for Use 63...
Page 64: ......
Page 72: ......
Page 106: ...VTI Instruments Corp 106 EX1200A Triggering...
Page 108: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL...
Page 142: ...VTI Instruments Corp 142 EX1200A Web Page Operation...
Page 144: ...VTI Instruments Corp 144 EX1200A Series Programming...
Page 154: ......
Page 160: ......