EX1200A Triggering
99
R
0
= nominal resistance of RTD (resistance at 0 °C)
T
= temperature in °C
A
= α[1+ (δ/100)]
B
= -αδ/1 x 10
4
C
= -αβ/1 x 10
8
)
For the RTD’s with built-in support, the following Callendar-Van Dusen RTD parameters are
used:
RTD Type
Standard
α
β
δ
R
0
PT100
ITS-90
0.003850
0.10863
1.49990
100 Ω
D100
ITS-90
0.003920
0.10630
1.49710
100 Ω
F100
ITS-90
0.003900
0.11000
1.49589
100 Ω
PT385
IPTS-68
0.003850
0.11100
1.50700
385 Ω
PT3916
IPTS-68
0.003916
0.11600
1.50594
3916 Ω
2-Wire Measurement Connections
A 2-wire RTD measurement is connected in the same manner as a 2-wire resistance measurement.
The HI and LO terminals of the DMM are connected to the RTD’s two terminals (no polarity
restrictions).
DMM Front Panel
Sense HI
Input HI
Input LO
Sense LO
RTD
F
IGURE
4-15:
RTD
2-W
IRE
C
ONNECTION TO
DMM
F
RONT
P
ANEL
When using the DMM bus backplane through a switch multiplexer, as shown in Figure 4-16,
observe the common practices introduced in the voltage and resistance measurement discussions
above. Shielding greatly reduces the effect of external interferences/noises. It should also be noted
that the shields should only be connected to ground at one end of the cable and not both.
Summary of Contents for EX1200A SERIES
Page 32: ......
Page 50: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use...
Page 63: ...www vtiinstruments com EX1200A Series Preparation for Use 63...
Page 64: ......
Page 72: ......
Page 106: ...VTI Instruments Corp 106 EX1200A Triggering...
Page 108: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL...
Page 142: ...VTI Instruments Corp 142 EX1200A Web Page Operation...
Page 144: ...VTI Instruments Corp 144 EX1200A Series Programming...
Page 154: ......
Page 160: ......