EX1200A Web Page Operation
123
System Buttons
To the right of the
Status Display
are two system buttons that affect the functionality of the
EX1200A system and all installed modules.
Lock/Unlock
: Requests (or releases) exclusive access to the EX1200A system.
Reset
: Resets the SFP and the connected modules to power-on defaults.
Self Test
: Performs a self-test operation of the DMM hardware. See
more information.
Basic Settings Field
The
Basic Settings
field allows the user to select the type of measurement that will be made by
the DMM as well as specifying simple measurement parameters.
Measurement Mode
buttons: These buttons control the type of measurement the DMM will
make.
o
V--
: Sets the DMM to take dc voltage measurements (DCV).
o
V~
: Sets the DMM to take ac voltage measurements (ACV).
o
Hz
: Sets the DMM to take frequency measurements.
o
A--
: Sets the DMM to take dc current measurements (DCI).
o
A~
: Sets the DMM to take ac current measurements (ACI).
o
1/Hz
: Sets the DMM to take period measurements.
o
Ω 2W
: Sets the DMM to take 2-wire resistance measurements.
o
Ω 4W
: Sets the DMM to take 4-wire resistance measurements.
o
°C
: Sets the DMM to take temperature measurements.
Range
: Indicates the input range the DMM will use when a measurement is taken. The drop
menu provides different ranges depending on the
Measurement Mode
that was selected.
o
For
DCV
and
ACV
the following ranges are available:
Auto Range On
,
Auto Range
Off
,
Auto Range Once
,
100 mV
,
1 V
,
10 V
,
100 V
, and
300 V
.
o
For
DCI
and
ACI
the following ranges are available:
Auto Range On
,
Auto Range Off
,
10 mA
,
100 mV
,
1 A
, and
3 A
.
o
Hz
measurements are performed at
1.5 MHz
range, while
1/Hz
measurements are at
0.3333
s.
o
For
Ω 2W
and
Ω 4W
measurements, the ranges are:
Auto Range On
,
Auto Range Off
,
10
Ω,
100
Ω,
1 k
Ω,
10 k
Ω,
100 kΩ
,
1 MΩ
,
10 MΩ
, and
100 MΩ.
NOTE
When
Auto Range
is selected, all measurements dynamically determine the most appropriate
range. Selecting
Auto Range Off
causes all future measurements to use the last actual range. The
Auto Range Once
function is equivalent to using
Auto Range On
for one measurement, then
switching to
Auto Range Off
for the subsequent measurements.
Option
: Provides various options depending on the
Measurement Mode
selected above.
o
For
DCV
: Provides a
10 M
(Ω) or
HIZ
option. The
10 M
option should be used to
minimize noise pickup, while
HIZ
should to reduce loading errors.
o
For
ACV, ACI, Hz,
and
1/Hz
: Provides
High Filter
,
Medium Filter
, and
Low
Filter
options. The
High
option utilizes the fastest, but least accurate, ac filter. The
Low
option
selects slowest, most accurate ac filter. The
Medium
Filter
option provides a
compromise between the speed and accuracy of the other two filters.
o
For
DCI
: No options are provided.
o
For
Ω2W & Ω4W
: Provides the
HI_I
and
LO_I
options. The
HI_I
option uses the high
test current value, while the
LO_I
uses the low current value. Only the 100 Ω, 10 kΩ,
1 MΩ, and 10 MΩ ranges offer two current sources; this option will have no effect on
other ranges.
Input
: Indicates whether the input is from an
Internal
source or from the banana jacks on the
DMM’s
Front Panel
.
Summary of Contents for EX1200A SERIES
Page 32: ......
Page 50: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use...
Page 63: ...www vtiinstruments com EX1200A Series Preparation for Use 63...
Page 64: ......
Page 72: ......
Page 106: ...VTI Instruments Corp 106 EX1200A Triggering...
Page 108: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL...
Page 142: ...VTI Instruments Corp 142 EX1200A Web Page Operation...
Page 144: ...VTI Instruments Corp 144 EX1200A Series Programming...
Page 154: ......
Page 160: ......