Calibration Procedure
1)
Connect to the embedded web page. See
2)
Click
Soft Front Panel
from the
Index
page to navigate to the
Soft Front Panel
page. Select
the
EX1200-dmm DMM
from the list of available plug-in modules (the slot designation will
vary depending on the mainframe being used).
F
IGURE
9-6:
I
NSTRUMENT
SFP
S
ELECTION
P
AGE
3)
At the top of the DMM’s soft front panel, click the
Calibration
button.
F
IGURE
9-7:
C
ALIBRATION
T
AB
4)
From the
Calibration Functions
section of the
Calibration
page, select
Calibration
from
the drop down menu.
F
IGURE
9-8:
C
ALIBRATION
F
UNCTION
S
ELECTION
Select the DMM
Calibration Tab