Operating Instructions - CPC3050
77
High-Speed Pressure Controller
CPC3050
Table 7.6.1 - CPC3050 SCPI Command Set
Command
Response/Function
STATus
:OPERation
:CONDition?
Returns an integer value representing instrument
status that can be decoded.
Bit 0: Zeroing active
Bit 1: Control Setpoint has not been reached.
Bit 2: Reserved 0.
Bit 3: Reserved 0.
Bit 5: Measuring. The instrument is actively mea-
suring
MEASure
[:PRESsure][R]?
Returns the pressure from range R
:TEMPerature[R]?
Returns the temperature from range R
:RATE[R]?
Returns the rate/sec from range R
:BAROmetric?
Returns the barometric pressure
CALibration
[:PRESsure][R]
:MODE?
Returns 1=calibrated or 0=not calibrated
:DATE?
Returns date of cal “MM/DD/YY”
:DATE <i,i,i>
Sets date of cal YYYY,MM,DD
:ZERO?
Returns zero offset
:ZERO <n>
Sets the zero offset
:ZERO:RUN
Start Autozero sequence
:ZERO:STOP
Stop Autozero sequence
:ZERO:INITiate?
Returns zero status
:ZERO:INITiate
Ignored
SENSe
[:PRESsure][R]
:NAME?
Returns transducer name string
:MODE?
Returns “ABSOLUTE” or “GAUGE”
:MODE ABS | GAUGE
Sets pressure type
:ABS?
Returns native transducer type
0=GAUGE 1=ABSOLUTE
:RESolution?
Returns resolution (float)
:RANGe
[:UPPer]?
Returns maximum range
:LOWer?
Returns minimum range
:UNIT
[:NAME]?
Returns ASCII units (mixed case)