88
Operating Instructions - CPC3050
High-Speed Pressure Controller
CPC3050
:VERSion[n]?
Returns the software version of the device N in
double quotes. When N is not specified, default
is 1.
Ex// INST:VERS2?
:INST:VERS “v1.2”
N
Device
1
Instrument Software
2
Operating System (OS)
3
4
Regulator PIC
5
Regulator FPGA
6
7
8
Sensor 1
9
Sensor 2
10
11
12
Barometer
13
GPIB
Ex// INST:VERS?
:INST:VERS “v2.27.0“
OUTPut[x]
[x]=x=1=Channel A
[:ISOLation]
:STATe?
Returns 1 if in Measure mode, 0 otherwise
:STATe ON | OFF | 1 | 0
ON or 1 puts in Measure mode, 0 puts in Vent
mode
:MODE Measure | Control | Vent
Sets the mode indicated. May use as low as 1
character in the data, i.e. ‘V’ for Vent, or ‘VENT’
for vent, etc.
:MODE?
Returns the mode string
:STABle?
Returns 1 if in Control mode and pressure is
stable, 0 otherwise
:STATe ON | OFF | 1 | 0
ON or 1 puts in Control mode , OFF or 0 puts in
Measure mode
[:STATe]?
Returns 1 if in Control mode, 0 otherwise
SENSe[x]
[x]=x=1=Channel A