64
Operating Instructions - CPC3050
High-Speed Pressure Controller
CPC3050
Acquire?
15 char string.
Ex:
Acquire? Test_stand_1
Returns:
<sp>(YES or NO), CCC…
CCC<cr><lf>
This command is used when multiple
computers would like to control the
instrument.
Yes if acquisition is successful
No if instrument is being controlled
with another computer
CCC… = name of controlling com-
puter
(see Release? and Unlock)
Address
1-31
Sets the GPIB Address
Address?
<sp>nn<cr><lf>
Returns the GPIB Address
Asset_tag
16 char string
General purpose string for customer
use
Asset_tag?
<sp>ssssssssssssssss<cr><lf>
Return customer asset tag string
Autorange
ON or OFF
Sets whether the autorange function
is enabled or disabled
Autorange?
<sp>(ON or OFF)<cr><lf>
Returns whether the autorange func-
tion is enabled or disabled.
Autozero
none
Re-zero all the ranges on the active
channel. These adjustments are not
password protected and are
saved
permanently through power cycles
while the instrument is in native mode
(temporary while in emulation mode)
.
This command takes approximately
60 seconds but may take longer de-
pending on the time to become stable.
Autozero?
S,T,X,X
Returns autozero data where S
represents state (0 = complete, 1 =
local autozero, 2 = remote autozero),
T represents the estimated time to
complete in seconds, and x is a (0)
character since this data location is
not used at this time.
Autozeroabort
none
Aborts autozero. Any transducers that
have been zeroed will not revert to
previous zero offsets.
Auxdisp <n>
NONE,PEAK,RATE,TEMPERATURE,
UNCERTAINTY,UNITS
Sets the auxiliary display setting for
index <n> (0 to 2). If <n> is left out,
defaults to index 0.
Auxdisp? <n>
<sp>NONE,PEAK,RATE,TEMPERAT
URE,UNCERTAINTY,UNITS<cr><lf>
Returns the auxiliary display setting
for index <n>. If <n> is left out, de-
faults to index 0.
Baro?
<sp>+n.nn<cr><lf>
Returns reading from barometric
transducer or “NO BAROMETER” if
one isn’t installed
Barounits
units code or text in table below
Sets the barometer pressure units