Operating Instructions - CPC3050
73
High-Speed Pressure Controller
CPC3050
Srqmask?
<sp>{string}<cr><lf>
Returns “stable”, ”error” or “error,
stable” depending on the SRQ.
Sstop
1 or 2
Sets the serial and USB stop bits
Sstop?
<sp>X<cr><lf>
Returns the serial and USB stop bits
Stable?
<sp>(Yes or No)<cr><lf>
Returns YES if instrument is stable or
No
Stabledelay
0 to 3600
Sets the stable time to the number of
seconds specified
Stabledelay?
<sp>XXXXXXX<cr><lf>
Returns the stable time.
Stabletime
0 to 3600
Sets the stable time to the number of
seconds specified
Stabletime?
<sp>XXXXXXX<cr><lf>
Returns the stable time.
StableWin
%FS value
Sets the stable window as a %FS
StableWin?
<sp>n.nn<cr>
<lf>
Returns the stable window.
Step
Value inside upper and lower limits
and inside the range of the active
transducer.
Sets the control step size for the
instrument
Step-
Jogs the setpoint down one step
Step+
Jogs the setpoint up one step
Step?
<sp>+n.nn<cr>
<lf>
Returns the control step for the instru-
ment
Steppct
Value in % of current range
Sets the control step in % of current
range
Steppct?
<sp>n.nn<cr>
<lf>
Returns the current step in % of cur-
rent range
Supply_Detect
YES or NO
Detects whether enough supply
pressure is connected, see Section
6.4.3.8 Detection Flags
Supply_Detect?
<sp>(YES or NO)<cr><lf>
Returns whether detection is enabled
Tare
ON / OFF
Tares the reading to zero
Tare?
<sp> n.nn <cr><lf>
Returns value of Tare
Temp?
Read temperature
TEMPUNITS?
Read temperature units (C, F, or K)
TEMPUNITS
<sp>(C, F, or K)<cr><lf>
Sets the temperature units to Celsius,
Fahrenheit, or Kelvin.
Termchar
CCCC
Set the output termination
character(s). CR, LF, CRLF, DEFAULT,
EOI, NONE
Termchar?
Returns the termination character
setting
Touchcal
Initiates a touchscreen calibration.
Transfer_factory_to_
linearity
Copy factory linearity coefficients to
customer
Units
units code or text in table below
Sets the instrument engineering units