42
Operating Instructions - CPC3050
High-Speed Pressure Controller
CPC3050
6.4.3.1 Control Behavior
The Control Behavior button in the Control Settings App provides a choice between different levels of
control behavior ranging from “0” to “100” where “0” represents lowest overshoot control mode and “100”
represents High Speed control mode. The Control Behavior is preset to “50” for Industrial control mode
and to “100” for High Speed control mode. The Control Behavior can be changed using the sliding scale.
Figure 6.4.3.1 shows the sliding bar to adjust control behavior in the Custom control mode.
Figure 6.4.3.1 – Control Limit Setting
6.4.3.2 Stability Parameters
Stability parameters for the controlled pressure can be found in the Control Settings App and can be
configured using the Stable Window and Stable Delay buttons. When the controller enters a stable condi
-
tion the pressure indication color on the Home App will change from white to green. The Stable Window
button allows the user to enter a value as a percentage of the highest range transducer. This value repre-
sents the pressure window within which any setpoint value would be considered stable by the user. The
Stable Delay button lets the user add a desired delay until the pressure value is considered stable while
being in the stable window.
Stability
Parameters
Figure 6.4.3.2 – Stability Parameters