80
Operating Instructions - CPC3050
High-Speed Pressure Controller
CPC3050
:TOLerance<n>
Sets the stable window
CALCulate
:LIMit
:LOWer<n>
Set the minimum control limit
:LOWer?
Set the minimum control limit
:UPPer<n>
Set the maximum control limit
:UPPer?
Set the maximum control limit
7.6.2 Error Messages
All remote commands sent to the CPC3050 are shown in the Trouble Shooting App, under the “Remote”
button. If there is a syntax error, an error message will appear below the errant command. Local errors
and remote command errors will appear under the “Error” button. Maximum of 100 errors are stored and
can be retrieved.
The error messages can also be seen remotely when “SYST:ERR?” is sent any error in the error queue is
returned, otherwise the following is returned: 0,“No Error”.
Error String Returned
0, “No error”
1,”Overflow”
102, “Syntax error”
114,”Parameter Out Of Range”
500,”<s>” (where <s> is the undefined error error string generated by the pneumatic module)
602,”Sensor not available”
999,”<s>” (where <s> is the undefined (unique) error string)