Operating Instructions - CPC3050
High-Speed Pressure Controller
CPC3050
6.4.2.6
Configuration
Configuration is the last parameter in the Settings App. It allows the operator to save instrument settings
and load them, as needed, in the future. Parameters that are set in the Settings App, Display App, and
the Control Settings App can be saved using the Configuration “Save” button and recalled using the Con
-
figuration “Load” button. Simply set all desired parameters then go to Settings-Configuration, press one
of the numbered Configuration buttons then press the “Save” button. This will save the current configura
-
tion in that button. To reload a saved configuration at a later time, go to Settings-Configuration and press
the numbered configuration button corresponding to the saved configuration and then press the “Load”
button.
Figure 6.4.2.6 – Configuration
The instrument default configuration can be activated simply by pressing the “Default” Button.
Parameters saved:
Language, User defined units and multipliers, volume, brightness, bar graph, auxiliary displays, cal func
-
tion, pressure type, filter level, resolution, rate units, temperature units, upper/lower control limit, step
size, stable delay, stable window, vent limit, control volume setting, control mode (High Speed, Industrial,
Custom), control behavior, command set, and termination character.