Barounits?
<sp>CCCC<cr><lf>
Returns the barometer units in a text
string
Beep
YES or NO
Sets the speaker volume to On or
Off, where On = 100%, and Off = 0%.
Related to command LOUDNESS.
Beep?
<sp>(YES or NO)<cr><lf>
Returns if the speaker volume is non-
zero.
Brightness
0 to 100
Set the brightness of the screen from
0 to 100%.
Brightness?
<sp>nn<cr><lf>
Returns the brightness of screen.
Calculate_as_found_
linearity
Calculate linearity slopes and inter-
cepts from true/actual pressures
Caldisable
YES,NO
Sets whether or not calibration of the
active transducer is disabled.
Caldisable?
<sp>(YES or NO)<cr><lf>
Returns whether or not calibration of
the active transducer is disabled.
Calibration_mode?
<sp>(YES or NO)<cr><lf>
Returns the current calibration mode
setting
Calibration_mode
YES, NO
Enable/ disable the calibration mode,
which when on, disables the sensor
overpressure error and forces range
hold to lock the secondary shutoff
valve. Normally the secondary range
is always allowed to go to its maxi-
mum pressure to avoid a large pres-
sure disturbance when switching to it.
Cerr
None
Clears the error queue
Chan
A or Baro
Set the active remote channel. Chan-
nel Baro should only be set when
used for calibration functionality
Chan?
<sp>A or BARO<cr><lf>
Returns currently active remote chan-
nel.
Cmdset
Mensor, DPI510, SCPI
Activates remote command set for
instrument emulation modes.
Cmdset?
<sp><CCCCCC><cr><lf>
Returns active command set identifier
Configuration?
<sp>INSTRUMENT: MENSOR,
CPC3050, 41xxxxxx, x.xx.xx<cr><lf>
<GPIB version><cr><lf>
CHANNEL A: <REG A ID
String><FPGA version><cr><lf>
<Channel A Primary sensor ID
String><cr><lf>
<Channel A Secondary sensor ID
String><cr><lf>
Returns the serial number and version
of the instrument, serial number and
versions of each channel’s regulator
PIC and FPGA, each sensor’s serial
number and version, and a list of all
enabled options
Control
Instrument placed in Control Mode