Menicon ShellPa Pro Instruction Manual Download Page 7

SP_Inst02 /02_20220208 

Instruction Manual For ShellPa Pro (Model type : STMP) 

Page(7/32) 

Lifescience  Dept.  Menicon  Co.,Ltd. 

1.4. Switch on and check initial position (original point). 

Below procedure, please check and conduct. 

 

Beforehand, please check nothing of an obstacle in chamber holder setting area on 

Operation unit

.-

-- [1] 

 

 

Please turn on a 

main switch

 

on “

C

ontrol unit”

, if nothing. --- [2] 

Initial set up movement on “Operation unit” start.

 (start to move in order to return to 

origin.)

 

After finishing to return to origin, normally, 

“Operation unit”

 will stop, a

nd “

Main menu

 screen on 

“Main menu” of “

Control unit

”,

 will be displayed, automatically. --- [3] 

 

       

       

 

[1] Check nothing of obstacles                  [2] Switch on                      Demo screen (automatically change) 

 

[3] 

“Main menu” screen

 

 

Summary of Contents for ShellPa Pro

Page 1: ...SP_Inst02 02_20220208 Instruction Manual For ShellPa Pro Model type STMP Page 1 32 Lifescience Dept Menicon Co Ltd Instruction Manual For ShellPa Pro Model type STMP...

Page 2: ...stretch and relax parameter P13 Setting Stretch ratio Setting Stretch and relax speed time 3 3 Operating to start and stop P15 Start operation Stop operation Forced Stop 4 Cycle Stop system mode P17 4...

Page 3: ...Hour Meter P25 7 3 Set Position and Pin Gap Adjust P25 8 Maintenance and Overhaul alarm P27 Maintenance When operating time reaches 2 000 hours Overhaul When operating time reaches 8 000 hours 9 Alarm...

Page 4: ...ipment 1 1 Checking the supplied items First please check there are below units and parts Operation unit 1unit Control unit 1 unit Stretch chamber holder 1holder Cover plate 5plates Stretch chamber SC...

Page 5: ...eration unit Control unit Stretch chamber holder Connecting cable Stretch and relax point bar Rigid point bar Access cover Power cable connector Main switch Main display Fuse Connecting cable Vent hol...

Page 6: ...gure1 Connecting and setting ShellPa Pro Beforehand cleaning up closely inside and outside of operation unit and stretch chamber holder using ethanol After install it into incubator Like figure1 pleas...

Page 7: ...tacle in chamber holder setting area on Operation unit 1 Please turn on a main switch on Control unit if nothing 2 Initial set up movement on Operation unit start start to move in order to return to o...

Page 8: ...ain menu H C Stretch Stop system Setting Parameters H C Operating Pattern A Operating Pattern B1 Operating Pattern B2 Operating Pattern C Same layout for PatternB1 Same layout for PatternC Same layout...

Page 9: ...ng Pattern A Setting Setting parameter of stretch and relax movement in Pattern A Data Output Output data parameter wave record etc of stretch and relax movement in Pattern A Stretch Stop system Monit...

Page 10: ...chamber Photo Cell seeding on and cell attaching to Stretch chamber in Menicon Co Ltd Cell L929 Chamber SC4Dea Check time 24h after seeding You can use SC4Dea or SC4Ha We recommend quantity of cultur...

Page 11: ...Please check below explanation of each pattern Pattern A Sq Tri ST Square wave Standstill You can practice stretch and relax cell culture on constant speed And you can set standstill time Please refe...

Page 12: ...t standstill time in addition to Pattern B1 If retention time set on zero movement will be the same as PattenB1 Please refer to wave on Operating Pattern screen below Pattern C Twice square wave Stand...

Page 13: ...is 20mm Stretch ratio 1 1 Stretch length is 0 2mm Stretch ratio 20 Stretch length is 4 0mm Setting Stretch and relax speed time Here we explain about Stretch and relax speed time on a case of pattern...

Page 14: ...ameter on Pattern B2 Parameter sign Contents Input area A One cycle time Cannot input data B Stretch movement time to specified stretch ratio length 0 25 30 00sec C Retention time on stretch 0 0 sec 8...

Page 15: ...initial position 0 0 sec 86400 0sec 24h 3 3 Operating to start and stop Start operation 1 Chose which pattern of stretch on main menu screen Display will go to Operating Pattern screen you had chosen...

Page 16: ...tention START button is flashing and STOP button is lighting when machine completely stop And SET Position is displayed and light same its around area SET Position indicates that stretch and relax poi...

Page 17: ...mode Setting On Main menu screen please set off Stretch Stop system because Cycle Stop system and Stretch stop system cannot be used at the same time Next go to Operating Pattern setting screen on eac...

Page 18: ...stretch and relax cycle until stop as Remaining Count on Operating Pattern screen is displayed Remaining Count on Operating Pattern setting screen is displayed in Remaining Count on Operating Pattern...

Page 19: ...retch Stop system mode ON OFF Stretch Stop System ON Stretch Stop System ON sign and button light on Stretch Stop System OFF Stretch Stop System OFF sign and button light off 5 2 Stretch Stop system m...

Page 20: ...cycle stretch and relax movement Stretch Stop system changes from Stretch Time mode to Stop Time mode Therefore there is possibility to take Stretch Time take time longer than setting time refer to e...

Page 21: ...ystem Number of sets is full below screen will be displayed on Operating Pattern screen and RESET button on Stretch Stop system Monitoring screen and is flashing And Stretch and relax movement stops A...

Page 22: ...ard or taking out and in SD card clear reset bar graph because this bar graph directly does not check SD card SD card insert On case of no SD card Memory No Set is displayed on Main menu screen SD car...

Page 23: ...cond after it change over from SD card to USB First Store Data in SD card Please Check Stretch and relax wave data displayed on Operating Pattern data output screen If O K Please press and hold for 3...

Page 24: ...he display change from USB memory to SD card USB memory screen Next Select Display data on SD card USB memory screen Started to write data to USB and Not pull out USB is displayed When writing data on...

Page 25: ...Hour Meter You can check total operation time on this machine But you cannot change this number check only 7 3 Set Position and Pin Gap Adjust These parameters mean Set Position Position to set Chamb...

Page 26: ...tion Manual For ShellPa Pro Model type STMP Page 26 32 Lifescience Dept Menicon Co Ltd Stretch and relax movement Push Stop Pin Gap Adjust Set Position Stationary Attention Machine is always located o...

Page 27: ...e just information Therefore on these alarms equipment can operate ordinary Maintenance When operating time reaches 2 000 hours Maintenance alarm flashes on Main menu screen when operating time reache...

Page 28: ...il Alarm contents Please check below table6 Alarm sign Table6 Alarm sign Alarm sign Contents Memo MC20V Alarm Power Off Internal calculation error Controller equipment system have abnormality Please t...

Page 29: ...Check machine stops or STOP button lighting off on Operating pattern screen You can change Operating pattern screen to Main menu screen when Stretch and relax movement completely stops It is interlock...

Page 30: ...m H155 Connecting cable length 2 3m Mass Approx Operation unit 5 8 kg Control unit 3 7kg Power requirement 100 to 240V 47 to 63Hz Output Voltage D C 24V 2 5A 60W Stretch ratio Input ratio range 1 to 2...

Page 31: ...he dish to completely cover the bottom of the chamber 4 Incubate the dishes for 30 minutes to 4 hours at 37 C If the coating is not sufficient please extent the incubation time 5 Remove the coating so...

Page 32: ...add 1 to 1 5 mL of the diluted PDL solution 2 to the chamber Gently tap the dish to completely cover the bottom of the chamber 4 Incubate the dishes for 1 to 4 hours at 37 C If the coating is not suff...

Reviews: