SP_Inst02 /02_20220208
Instruction Manual For ShellPa Pro (Model type : STMP)
Page(13/32)
Lifescience Dept. Menicon Co.,Ltd.
3.2. Setting stretch and relax parameter
You can set parameter on
“Operating Pattern setting” screen
.
Each “Operating Pattern setting” screen
is not same on each pattern. But standard parameters is
two parameters below.
(1)
Stretch ratio
(2)
Stretch and relax speed (time)
<Setting Stretch ratio>
Please touch the white part next to “1=” on “Operating Pattern setting” screen
.
After, “Key pad” will
be displayed on the screen. You can enter
“Stretch rat
io
”
. Parameter which you can enter is from 1%
to 20%.
Stretch ratio 100% is 20mm,
Stretch ratio 1=1%
---------- Stretch length is 0.2mm.
Stretch ratio=20%
---------- Stretch length is 4.0mm.
< Setting Stretch and relax speed (time)>
Here, we explain about Stretch and relax speed (time), on
a case of “pattern A”.
Please touch the white part next to each parameters (B, C, D, E)
on “Operating Pattern setting”
screen
. After, “Key pad” will be displayed on
the screen. Please refer to table1. about numbers which
you can enter, and contents of each parameter.
Tabel1. Each parameter on
“Pattern A”
Parameter sign
Contents
Input area
A=
One cycle time
Cannot input data
B=
Stretch movement time to specified
stretch ratio (length)
0.25 --- 30.00sec.
C=
Standstill time on stretch
0.0 sec. --- 86400.0sec.(24h.)
D=
Relax movement time to non-stretch
initial position
0.25 --- 30.00sec.
E=
Standstill time on relax (non-stretch
initial position)
0.0 sec. --- 86400.0sec.(24h.)