9.2. Trouble shooting
If you have any problems, please check below Table7.
Table7. Trouble shooting
Trouble
Check
Memo
Electricity is not
supplied
Check connecting
“
Power supply
adaptor
”
,
“
Power supply cable
” and
“Control unit”.
---
Operation unit do not
move.
Check connecting “
Connecting
cable
” and “Control unit”.
---
Cannot operate to
change “Operating
pattern” screen to
“Main menu” screen.
Check machine stops
, or “STOP”
button lighting off
on “Operating
pattern” screen
You can change “Operating
pattern” screen to “Main menu”
screen, when Stretch and relax
movement completely stops.
(It is interlock.)
Stretch and relax
position changed.
Had you already conducted
operation of return to origin ,
removing “
Stretch chamber
holder”.
Please conduct operation of return
origin
, removing “Stretch chamber
holder”
.
(refer to “
1.4. Switch on and check
initial position (original point)
”)
O
n setting “Stretch chamber
holder”
, operating return to
origin, there is possibility to
change the position of origin.
And, It’s same, in case of
operating “Forced stop”. (refer to
<Forced stop> in “
3.3. Operating
to start and stop
”)