46. Click on the
Run Auto Test
button and follow the prompts provided by the tester for setting power pack
controls and for testing lead 1 then lead 2 for each coil.
47. When the
Surge Test
window opens, notice that the
Test-REF
and
EAR(%)
elements are active now that a test
reference has been set.
Fig 38:
Testing coils using the test reference.
48. When you have completed testing on both leads for a coil, you can close the Surge Test window and save the
test data collected.
49. Disconnect the coil from the tester.
NOTE: The software also provides features to save up to three waveforms in addition to the reference waveform for
reporting and comparison needs. For additional information on single-coil testing and details on related features,
refer to the
Baker AWA-IV User Guide
.
50. If you need to test another coil, connect it to the tester as described above and repeat the process starting from
step 42.
Using Power Packs with Baker AWA Testers
www.megger.com 41
Summary of Contents for PPX30