Creating Test and Motor IDs, and Setting Up the Tests
NOTE: The processes for creating Test IDs and Motor IDs is summarized in this guide. Their inclusion is intended
only to provide context for the overall process. Refer to the
Baker AWA-IV User Guide
for more detailed information
about each process and the related user interface elements.
Create a New Test ID
1. Click on the
Tests
tab then click on the
Edit Test ID
checkbox.
2. Enter the password for editing Test IDs.
Fig 13:
Tests tab, editing Test ID to create new Test ID.
3. Click on the
Add
button to open the
Create New Test ID
dialog box as shown below.
Fig 14:
Create New Test ID dialog box.
4. Click on the
Add Blank Test ID
radio button.
5. Enter the new Test ID in the appropriate field. In this example, we use PP30_6600.
6. Use the
Target Motor Voltage Class
drop-down menu to select a voltage class. For this example, we would
select a voltage class of 6600.
7. Click
OK
to continue.
Using Power Packs with Baker AWA Testers
22
Baker Power Packs User Guide 71-061RC EN V2
www.megger.com
Summary of Contents for PPX30
Page 1: ...Baker Power Packs PPX30 PPX30A PPX40 User Guide...
Page 2: ......
Page 6: ...Notices iv Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Page 12: ...Baker Power Pack Introduction 4 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Page 24: ...Baker Power Pack Setup 16 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Page 74: ...66 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Page 78: ...70 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Page 82: ...74 Baker Power Packs User Guide 71 061RC EN V2 www megger com...