Baker PPX40 Manual Surge Range Switch
The Baker PPX40 manual
Surge Range
switch for performing Surge and HiPot tests. This switch provides positive
control of the Surge test voltage and eliminates dependency on automatic range switching during a test.
During HiPot testing, set this switch in the 25-40K position. The Baker PPX does not limit the output voltage in HiPot
mode. The Hi/Low is for Surge testing. The LED still follows the switch, but the limit circuit is disabled.
Fig 9:
Manual surge range switch on PPX40.
Table 10:
Surge range switch positions and functions.
Switch Position
Test Voltage
Requirements
0–25kV Surge
< 25 kV
In this position, the voltage will be limited to 25kV.
25–40kV Surge
> 25 kV
In this position, voltage is allowed to increase above 25kV, but
surge pulses will not be generated if the voltage is lowered
below 25kV.
Baker Power Pack Introduction
www.megger.com 15