8
4.6.3 Electromagnetic compatibility – emissions & immunity
Guidance and manufacturer’s declaration – Electromagnetic emissions
The MEG-ENGINE II is intended for use in the electromagnetic environment specified below. The customer or the user of the MEG-ENGINE
II must ensure that it is actually used in such an environment.
Guidance and manufacturer’s declaration – Electromagnetic immunity
The MEG-ENGINE II is intended for use in the electromagnetic environment specified below. The customer or the user of the MEG-ENGINE
II must ensure that it is actually used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The MEG-ENGINE II uses RF energy for its internal operation only. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The MEG-ENGINE II is suitable for use in any building, including residential buildings
and those directly connected to the public low-voltage power supply network that sup-
plies buildings used for residential purposes.
Harmonic emissions IEC
61000-3-2
Class A
Emissions due to voltage
fluctuations IEC 61000-3-3
Conforming
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV air
±4 kV air
±8 kV air
±15 kV air
±8 kV contact
±2 kV air
±4 kV air
±8 kV air
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for other lines
±2 kV for power supply lines
N.A.
Mains power quality should
be that of a commercial or
hospital environment.
Surge
IEC 61000-4-5
±0.5 kV line to line
±1 kV line to line
±0.5 kV line to earth
±1 kV line to earth
±2 kV line to earth
±0.5 kV line to line
±1 kV line to line
±0.5 kV line to earth
±1 kV line to earth
±2 kV line to earth
Mains power quality should
be that of a commercial or
hospital environment.
Voltage dips, short inter-
ruptions and voltage varia-
tions on power supply
input lines
IEC 61000-4-11
0% U
T
for 0.5 cycle, at 0°, 45°, 90°,
135°, 180°, 225°, 270° and 315°
0% U
T
for 1 cycle and
70% U
T
for 25/30 cycles at 0°
0% U
T
for 250 cycles at 0°
0% U
T
for 0.5 cycle, at 0°, 45°, 90°,
135°, 180°, 225°, 270° and 315°
0% U
T
for 1 cycle and
70% U
T
for 25/30 cycles at 0°
0% U
T
for 250 cycles at 0°
Mains power quality should
be that of a commercial or
hospital environment. If the
user of the MEG-ENGINE II
requires continued opera-
tion during mains power
interruptions, it is recom-
mended that the MEG-
ENGINE II be powered from
an uninterruptible power
supply or a battery.
Magnetic field due to mains
frequency (50/60 Hz)
IEC 61000-4-8
30 A/m
30 A/m
Magnetic fields generated
by the mains frequency
should be at levels charac-
teristic of a typical location
in a typical commercial or
hospital environment.
Summary of Contents for MEG-ENGINE II
Page 27: ...ENG 25 NOTES NOTES ...
Page 28: ...26 ...
Page 53: ...FRA 51 REMARQUES REMARQUES ...
Page 54: ...52 ...
Page 79: ...DEU 77 BEMERKUNGEN BEMERKUNGEN ...
Page 80: ...78 ...
Page 105: ...ESP 103 NOTAS NOTAS ...
Page 106: ...104 ...
Page 131: ...ITA 129 Appunti Appunti ...
Page 132: ...130 ...
Page 157: ...PRT 155 NOTAS NOTAS ...
Page 158: ...156 ...
Page 185: ...RUS 183 ПРИМЕЧАНИЯ ПРИМЕЧАНИЯ ...
Page 186: ...184 ...
Page 196: ...194 图 1 图 2 图 3 图 4 1 2 1 1 2 图 5 图 6 图 7 图 8 1 2 AAAA MM JJ 1 2 3 图 9 图 10 图 11 5 安装 ...
Page 205: ...中文 203 图 3 图 4 1 1 注意 1 在特殊模式 操作脚控无效 2 经过所有特殊模式 方可再次显示设置模式 3 默认显示 reset no 文本 ...
Page 211: ...ZHO 209 注意 注意 ...
Page 212: ...210 ...
Page 222: ...220 図 1 図 2 図 3 図 4 1 2 1 1 2 図 5 図 6 図 7 図 8 1 2 AAAA MM JJ 1 2 3 図 9 図 10 図 11 5 セッティング ...
Page 237: ...JPN 235 注 注 ...
Page 238: ...236 ...
Page 263: ......