Masterflex
Portable Sampling Drive Operating Manual
5-2
E. Verify that the
Pump Rotational Selection On/Off Switch (H),
is either in the
CW
or
CCW
position for the desired rotation. The
Power
indicator in the
Drive Controls section will be illuminated.
F. Verify that the coaxial plug is fully inserted in the External Power Input of the
unit.
G. Verify that AC voltage is present in the AC outlet to which the AC/DC Power
Supply/Converter is connected.
H. Verify the DC voltage at the 2.5 mm coaxial connector. Must be at least 13
volts DC.
AC/DC Power Supply/Converter is not a serviceable unit and
contains no user serviceable parts.
I. If none of the above solutions work, contact your authorized dealer for
service.
3. The Drive will not operate using the Automotive Power Adapter Accessory.
A. Verify that the
Power Source Selection Switch (B),
is in the
EXTERNAL
position.
B. Verify that the
Mode of Operation Selection Switch (C),
is in the
RUN
position.
C. Verify that the pump head, tubing size and formulation are correct and
recommended for this drive.
D. If using a PTFE pump head, install and start the tubing per the instructions
in the pump head manual. Excessive tightening of the occlusion bed can
jam the drive mechanism. Tighten the occlusion bed gradually as indicated
in the pump head manual.
E. Verify that the
Pump Rotational Selection On/Off Switch (H),
is either in the
CW
or
CCW
position for the desired rotation. The
Power
indicator in the
Drive Controls section will be illuminated.
F. Verify that the coaxial plug of the Automotive Power Adapter is fully inserted
in the External Power Input of the drive.
G. Verify that the Automotive Power Adapter is fully inserted into the cigarette
lighter on the vehicle.
H. Verify that the fuse for the cigarette lighter is not open or blown.
I. Verify the DC voltage at the 2.5 mm coaxial connector. It must be at least 12
volts DC.
J. Verify the polarity of the DC voltage at the 2.5 mm connector. The sleeve
is negative, the center is positive.
The drive is protected against polarity
reversal and will not operate if the polarities are reversed.
K. If none of the above solutions work, contact your authorized dealer for
service.
Troubleshooting
(Continue)
Section 5