Operating Speed
70 to 400 rpm
Operating Voltage/Frequency:
Model 07571-02
115V AC nominal 50/60 Hz (100–130) V AC @ 120mA AC
230V AC nominal 50/60 Hz (200–260) V AC @ 60mA AC
12V DC nominal (11–16)V DC @2.4A DC
Number of Pump Heads:
1
Torque Load:
50 oz-in (0.35 N•m), maximum
Direction of Rotation:
Clockwise and Counterclockwise
Operating Temperature:
32°F to 122°F (0° to 50°C)
Storage Temperature:
−4°F to 149°F (−20°C to 65°C)
Humidity (Non-condensing):
10% to 90%
Altitude
6600 ft. (2000 m)
Dimensions (L x W x H):
17.9 in x 12.1 in x 9.6 in
(38 cm x 31 x 24 cm)
Weight:
17.8 pounds (8.07 kg)
Chemical Resistance:
High-density polyethylene, clear anodized aluminum,
stainless steel and polyester. All materials withstand
standard
solvents.
Enclosure Rating:
IP54 per IEC529
Pollution Degree:
Degree 2 per IEC664
Installation Category:
Category II per IEC664 (local level—appliances, portable
equipment,
etc.)
Compliance:
07571-02
115V AC Power Supply AC/DC Converter is UL listed and
CSA approved. Regulatory agency specifications not
applicable to the balance of the unit due to low voltage.
CE
Mark:
EN809 (EU Machinery Directive)
EU
EMC
Directive
(EN61326-1)