Portable Sampling Drive Operating Manual
Masterflex
D.
Adjust pump flow rate
with the 1-turn
Speed Control (I),
by turning the knob
CW to increase flow rate or CCW to decrease flow rate.
E. To
“Prime the Tubing”,
toggle the
“MAX” Switch (J),
up or down to make
the pump run at maximum speed.
To Operate Pump With an AC Power Source
DANGER:
NEVER apply AC voltages directly to the EXTERNAL POWER
INPUT receptacle on the front panel. The application of AC voltages
can result in injury and death of the operator and destruction of the
unit. Use only the AC/DC Power Supply/Converter supplied with the
unit to power unit from an AC source.
DANGER:
The AC/DC Power Supply/Converter is rated for INDOOR
USE ONLY. DO NOT use the AC/DC Power Supply/Converter in an
outdoor environment to either charge the battery or power the drive.
Electrical shock, severe injury and/or death is possible if this warning
is ignored.
A.
Connect the supplied AC/DC Power Supply/Converter to drive
by lifting the
EXTERNAL POWER INPUT (A),
receptacle cover and inserting the 2.5 mm
coaxial power jack into the receptacle. (See Figure 1.)
B.
Connect AC/DC Power Supply/Converter to an AC power source. Place
the Power Source Selection Switch (B), (EXTERNAL/BATTERY)
to the
“EXTERNAL”
position.
C.
Place the Mode of Operation Selection Switch (C) (CHARGE/RUN)
to the
“RUN”
position.
D.
Select Pump Rotational Direction Selection On/Off Switch (H),
CW or
CCW. (See Figure 1.) (When changing directions, turn off the power first,
which is the center or OFF position of the Rotational Direction Switch.)
E.
Adjust pump flow rate
with the 1-turn
Speed Control (I),
by turning the knob
CW to increase flow rate or CCW to decrease flow rate.
F.
To “Prime the Tubing”,
toggle the
“MAX” Switch (J),
up or down and hold to
make the pump run at maximum speed.
3-5
Drive Operation
(Continue)
Section 3