LED BRIGHTNESS
1.
With the monitor turned on, press the Brightness Button once to enter the LED
Brightness menu. The display will show the default setting Level 2.
2 Continue pressing the Brightness Button to scroll through the settings.
3.
Press the Enter Button to accept the desired setting and exit to the home display
screen.
TESTING THE RAD-87 WITH MASIMO SET TESTER (OPTIONAL)
1.
Turn the Rad-87 off and then on again.
2.
Connect the Masimo SET Tester to the Pulse CO-Oximeter Patient Cable Connector.
3.
Verify that within 20 seconds all available pulse bars display.
4.
Verify that the SpO
2
measurement is between 79% and 84%.
5.
Verify that the pulse rate measurement is between 55 bpm and 65 bpm.
6.
Set the SpO
2
low alarm limit to 90 (see Section 4,
Setup Menu Level 1, Parameter/
Measurement Alarm Limits - Screen 1,
and
Setup Menu Level 2, Alarm Volume
).
7.
Verify that an audible alarm activates, the SpO
2
measurement and the SpO2
parameter label are flashing, and the Alarm Bell and the System Status Light are
flashing red.
8.
Press the Alarm Silence Button once and verify that the alarm is silenced and the
Alarm Bell is flashing red and the System Status Light is flashing red.
9.
Wait 120 seconds and verify that the alarm silence times out, the audible alarm is
activated again and the Alarm Bell and System Status Light are flashing red.
10. Press the up arrow button several times and verify that the loudness of the pulse
beep tone increases.
11. Press the down arrow button and verify that the loudness of the pulse beep tone
decreases until the pulse beep tone is turned off.
12. Reset the device to original settings and remove the tester to complete the procedure.